Cannon miniAV-HT Bath 230V 50/60 HZ_1374520
Representative image only

Cannon miniAV-HT Bath 230V 50/60 HZ

Maximum thoughput12 tests per hour
Minimum sample/solvent volume5 mL sample/15 mL solvent per test (as little as 3 mL with fast run tubes)
Operating conditions15 °C to 30 °C, 10% to 75% relative humidity (non-condensing), Installation Category II; Pollution Degree 2
Temperature range and accuracy40 °C to 100 °C, ± 0.01 °C 100 °C to 150 °C, ± 0.02 °C * Alternate bath medium may be required for temperatures below 60 °C or above 135 °C
Timing Resolution0.01 s (timing accuracy to ± 0.001 s)
Viscosity Range0.5 mm2/s (cSt) to 10,000 mm2/s (cSt) in 100-fold increments (depending on viscometer tube selection).
miniAV-HT brochure
Overview

High Temperature Kinematic Viscometer

For Kinematic Viscosity of Asphalt Binders, Cements and Cutbacks

AASHTO T201, ASTM D2170, ASTM D445/D446, ISO 3104/3105, IP 71

miniAV-HT is a fully automated, single-sample, benchtop viscometer for determining the kinematic viscosity of asphalts at 60 °C and 135 °C. The Ubbelohde-style tube covers a 100-fold viscosity range at values between 0.5 mm²/s (cSt) and 10,000 mm²/s (cSt).

Meets AASHTO T201, ASTM D2170 and ASTM D445 precision requirements

  • Developed specifically for asphalt kinematic viscosity testing at 60 °C and 135 °C
  • Instrument temperature range: 40 °C to 150°C
  • Viscosity range: 0.5mm²/s (cSt) to 10,000mm²/s (cSt)* in 100-fold increments (easily covering the range of 5 separate manual glass viscometers)
  • A well tested CANNON viscometer platform offers reliability and dependable support
  • Includes adjustable sample preheater and heated waste lines

Simple, automated benchtop testing

  • Automation provides an alternative to labor-intensive manual testing and reduces operator to operator variability
  • VISCPRO® PC software enables one PC to manage and control up to four instruments
  • User-configurable reports may be viewed, printed, saved and exported
  • Up to 12 determinations/hour

Compact, convenient design

  • Modular bath for easy maintenance access
  • System fits in roughly the same benchtop area as an analytical balance

Reduced maintenance and consumable costs

  • Ubbelohde-style tube reduces solvent usage and disposal costs by 50%
  • Operators physically replace tubes in minutes, eliminating the need for related service calls
  • Single-point temperature calibration allows calibration at a single relevant test temperature and avoids the need for tube recalibration at other temperatures
  • Instrument includes standard dual-solvent input

* Some upper viscosity measurements may be limited by test temperature and sample type.