FLOWCONTROLR AL 15L/M AR GFC_1142696
Representative image only

FLOWCONTROLR AL 15L/M AR GFC

Accuracy±1.5% full-scale
CertifiedYes
Communicationslinear 0 to 5 VDC (1000 Ω minimum load), 4 to 20 mA (50 to 500 Ω loop resistance)
DescriptionThermal Gas Mass Flow Controller, to 15 L/min Ar, Aluminum body w/ display
Display Type3.5-digit LCD, detachable
Gas typeArgon
Height (Inch)6
Leak Integrity1 X 10-7 sccs He (max)
Length (Inch)1.25
Max flow rate (LPM)15
Max Operating Temperature (DegreeC)50
Max Operating Temperature (DegreeF)122
Max Particulate Size (microns)5
Max pressure (psi)1000
Media TypeArgon
Min flow rate (LPM)0
Min Operating Temperature (DegreeC)0
Min Operating Temperature (DegreeF)32
Output0 to 5 VDC (1000 Ω minimum); linear 4 to 20 mA (50 to 500 Ω loop resistance)
Output (Voltage)Yes
Power12 VDC, 1100 mA max
Pressure drop at maximum flow (PSID)3.87
Process connection1/4" Compression
Repeatability±0.5% full-scale
Response Time (msec)2000
Wetted MaterialsAnodized Aluminum, Brass, Viton
Width (Inch)5.25
View Virtual Catalog - English - gy44-0371.pdf
Overview
  • Precisely measure flow rates and totals
  • Proven design uses metal body components for added durability
  • Accumulated flow can be logged using an optional totalizer
  • Includes NIST-traceable certificate

These controllers feature an advanced straight-tube sensor that ensures accurate and repeatable results. Gas flow measurements are unaffected by moderate temperature and pressure variations at the inlet. The meter also allows a four-point calibration across the flow range to improve meter linearity.

The flow rate set point can be established by either a local potentiometer or by a remote 4 to 20 mA or 0 to 5 VDC signal. Output data is sent via a 0 to 5 VDC or 4 to 20 mA signal; an analog-to-RS converter is also available for data collection and analysis on your PC.

All controllers include a detachable 3-1/2-digit LCD that can be tilted up to 90° for easy reading. The display is remote mountable to three feet using extension cable 32662-70. The totalizer option 32650-22 will display total accumulated flow.

The controller is protected from polarity reversal or short circuit by a built-in resettable fuse. Aluminum models have wetted materials of anodized aluminum, brass, and Viton; stainless steel (SS) models have wetted materials of SS and Viton.

Accuracy:
These precalibrated flow controllers operate at inlet pressures between 5 and 60 psi and at gas temperatures between 59 and 77°F (15 to 25°C) while maintaining the stated percent full-scale accuracy and linearity. When operating beyond 5 to 60 psi, add ±0.01%/psi full-scale; if operating beyond 59 to 77°F (15 to 25°C), add ±0.15%/°C full-scale.

Whats Included:

NIST-traceable calibration report supplied by the manufacturer.

Over 40,000 Customers Trust John Morris